OK, I got the answer from Curt and thought I would report back:
As a work around, you should be able to use the ORP connector in pH mode. A pH reading above 7.0 is a negative ORP reading. For every 1 unit of pH change the ORP is changing by about 58 mV. An ORP reading of -200 is equivalent to a...
Do you think if I use the 700mv calibration solution and set it to 1000mv that I can use it as an accurate -300mv offset? Or does it not work like that?
HI Carlo, I am having the same problem with a neptune probe and a crossover cable I made. Did you ever remember what the workaround was to get it to read -redox ?
Thanks,
Dan